AFA4000/1

A pioneering airflow monitoring system developed to ensure the safety of users working with industrial and educational fume cupboards.

AFA4000/2

A pioneering airflow monitoring system developed to ensure the safety of users working with industrial and educational fume cupboards, featuring a unique customisable push button function.

AFA500

An entry level airflow monitor designed for applications where airflow velocity indication is not required. It is available with either a built in or remote unique airflow sensor that is used to measure face velocities or duct velocities and available fully flush (MK2) or semi flush (MK3).

AFA500/BSC

Airflow monitor with unique airflow sensor suitable for Bio Safety cabinets available fully flush (MK2) or semi flush (MK3).

AFA1000/1

A digital airflow monitor designed for applications where pushbutton functions are not required and is available with a range of remote unique airflow sensors used to measure face velocities or duct velocities. Available fully flush (MK2) or semi flush (MK3).

AFA1000/2

A digital airflow monitor designed for applications where pushbutton functions are required and is available with a range of remote unique airflow sensors used to measure face velocities or duct velocities. Available fully flush (MK2) or semi flush (MK3).

AFA1000/PR

A pressure monitor suitable for room or duct differential pressure monitoring and BSC pressure monitoring, available fully flush mounting (MK2).

AFA4000/E

A pioneering airflow controlling system developed to ensure the safety of users, working with industrial and educational fume cupboards and biological safety cabinets.

AFA1000/E

A digital airflow VAV controller available with a range of remote unique airflow sensors that are used to measure face velocities or duct velocities. It is available fully flush (MK2) or semi flush (MK3).

AFA1000/E Dual Output

It has a dual output allowing control of a secondary function such as Supply Air damper/inverter control or Fresh Air bleed damper control.

AFA3000/E

It has been specifically designed for Venturi Valve fume cupboard applications.

The controller will adjust the valve to the desired flow based on sash position. The mechanically pressure independent valve will accurately control and maintain design flows (as the system pressure/sash position changes). As the valve cannot detect changes in the face velocity of the fume cupboard , TEL incorporates one of their velocity sensors that will increase the flow of the valve if the face velocity drops below desired values.

AFA1000/RM

A wall mounted room differential pressure controller suitable for controlling the supply air into a room using a damper, valve or inverter.

AFA1000/BLD

A wall mounted fresh air bleed controller suitable for controlling the VAV system duct pressure using a fresh air bleed damper or fan inverter drive.

AFA1000/AHU

A wall mounted pressure controller suitable for controlling the duct pressure of an AHU using an inverter drive.

4NM Fumecupboard Actuator

Modulating damper actuator for adjusting dampers in technical building installations.

6NM SuperCap Actuator

Modulating SuperCap rotary actuator with emergency control function and extended functionalities for adjusting dampers in technical building installations and in laboratories.

ECON Damper

A precision engineered single blade butterfly damper designed for many years of maintenance free movement. The clear bodied design allows you to check for any debris caught on the internal blade.

Venturi Valve

A mechanically pressure independent control valve that provides precise and stable airflow.

AFA 5000

An intelligent touch screen room controller offering demand-control ventilation for up to 64 fume cupboards per room without the need for a building management system.

VAV
(Sash Position Control)

The controller operates using a floating set point dependent on sash position.

3 Term Control (Pushbutton Control)

The controller operates using fixed volume set points (Vmax/Vnorm/Vmin) independent of sash position.

VAV Box Control (Remote 0-10VDC Input control)

The controller operates from an analogue input voltage to a volume range between Vmin and Vmax parameter values.

CAV Box Control (Fixed Volume set point)

The controller operates to a fixed volume set point to give Constant Volume operation.

Auto Sash Controller

Although preliminary designed for operation on VAV fume cupboards, it can also be used on CAV fume cupboards for increased safety. It can be fitted to new or existing fume cupboards with a choice of three motor drive systems to suit each type of fume cupboard.

Dual Auto Sash Controller

A dual auto sash controller is also available to operate on fume cupboards with two sashes including walk in type, back to back type and side by side type sashes.

AFA4000

A pioneering airflow controlling and monitoring system developed to ensure the safety of users, working with industrial and educational fume cupboards and biological safety cabinets. The AFA4000 airflow monitor and controller is available with optional Auto Sash integration.

Rectangular VAV Terminal Units

The Variable Volume terminal type RVP-P enables precision control of the amount of air coming in to a room. It can be supplied with a variety of control options for different applications and is ideally suited for use with the TEL AFA5000 Room Space Controller.

Fitted from 2 x duct width to a T junction or a bend for reduced installation costs.

Circular VAV Terminal Units

The Variable Volume Circular type RVP-C provides volume flow control of extract and supply air systems across a wide range of sizes and volumes to suit any application. Ideally suited for use with the TEL AFA5000 Room Space Controller.

Fitted from 1 x duct diameter from a bend.

Rectangular CAV Terminal Units

Volume flow controllers type VRRK are used for maintaining a constant volume on extract and supply air systems covering a wide range of sizes and volumes to suit any application. Their task is to keep a default nominal value of the volume flow constantly, lastingly and independently of the varying pressure in the duct.

Constant volume flow controllers work with a smooth-running, asymmetrically angled plate, which guarantees a sensitive responding behaviour even in case of low volume flows.

Adjustable in the field over a set volume range, the Constant Volume Terminal type VRRK ensures a constant supply of air of up to 10,000m3/hr to a room.

Circular CAV Terminal Units

The volume flow controllers VRK are used in complex piping systems for automatically controlling the amount of air distribution. Their task is to maintain a predetermined desired value of the air flow for the supply or exhaust air of a room sustainable and independent from fluctuating channel air pressure.

Maintenance free and with a range of between 75 to 1,650m3/hr, the Constant Volume Terminal type VRK can be tailored to meet your specific air supply and extraction requirements.

The CAV units are field adjustable over a set volume range.

Rectangular Attenuators Type PZX

Reduce sounds and save energy with aerodynamically shaped, flame retardant sound attenuation splitters. Available in in 100, 200 and 300mm thicknesses in five standard cross-section dimensions.

Circular Attenuator Type PZM

For sound attenuation in ventilated ducts and ventilated areas.

Circular Attenuator Type PZC – with POD

For sound attenuation in ventilated ducts and ventilated areas.

LEV Monitor

TEL’s LEV monitor provides you with the reassurance that the flow-rate is maintained, protecting your employees and alerting operators of any problems immediately.

Designed specifically for low and high pressure, TEL’s LEV Monitor can be fitted to any ducting or LEV extraction device, to check the airflow is adequate and keeping your employees continually safe in their working environment.

Kitchen Hood VAV System

Available for single or multiple hood controls.

Gas Interlocking System

TEL’s kitchen hood VAV system can be supplied with an integrated Gas Interlock system, which will cut off gas flow in the event of kitchen ventilation failure. This is now a mandatory requirement in most commercial kitchens.

Flameproof Systems

TEL can provide Airflow Monitors and Airflow Controls suitable for Zone 1 and Zone 2 hazardous or flameproof areas, Gas group IIC, Temperature class T6.

For installations where the hazardous area is contained within the Fume Cupboard chamber the standard airflow monitor or controller can be used with certified zener barriers fitted to the airflow sensor.

For installations where any part of the laboratory is classified as the hazardous area the airflow monitors or controls are mounted in a Safe Area control panel fitted in a designated safe area (e.g. corridor) and remote Ex rated Indicators and alarms fitted to the Fume Cupboard. e.g.